| Abstract |
|---|
Key Words: Clinical trials Systematic review Therapeutic ultrasound
| Introduction |
|---|
|
|
|---|
We used a systematic review to examine whether there is sufficient evidence to accept the premise that therapeutic ultrasound is effective. That is, does active ultrasound used alone, or with other interventions, produce a different outcome than placebo ultrasound applied under the same conditions? Based on the available evidence, we also examined issues related to dosage and usage, specifically, the total energy and the energy density applied.
| Reviews of Ultrasound Research |
|---|
|
|
|---|
Based on their meta-analysis of trials of physical therapy treatments for soft tissue lesions of the shoulder, van der Heijden et al13 concluded that ultrasound is not effective and its use should be discouraged. This conclusion was later criticized as being based on methodologically inadequate research studies.14 The criticisms emphasized the need for studies using double blinding, an internally valid method of placebo treatment, and adequate group sizes and including details of the dosage of ultrasound used. That is, the type of study required for credible conclusions in this context are randomized controlled trials (RCTs), which also provide adequate dosage details.
In a recent systematic review, van der Windt et al15 analyzed 38 RCTs and controlled clinical trials (CCTs) (trials in which subjects were not randomly assigned to groups) of the effectiveness of ultrasound used to treat people with musculoskeletal disorders. They concluded that there is little evidence to support the use of active ultrasound therapy for treating people with those disorders. Most methodologically adequate studies (n=13) they reviewed lacked evidence of either meaningful outcomes or statistically significant differences from using ultrasound.
We have since identified additional RCTs of ultrasound published in the English language and not reviewed by van der Windt et al.15 Some of these RCTs, which were beyond the scope of van der Windt and colleagues' study, focused on tissue healing1622 or pain.23 Other trials focused on using ultrasound for either a combination of pain and soft tissue healing or to change a consequential functional loss.2433 In all trials, at least active and placebo ultrasound were applied to treat one of these conditions. In some trials, additional, identical interventions were given to both groups but in such a manner that any contribution of active ultrasound to patient outcomes was evident with analysis.
Given the importance of ultrasound in physical therapy, we believe that a new systematic review of ultrasound for pain and soft tissue is necessary. Besides additional studies being available, we will use criteria that are slightly different from those used in the reviews discussed23,5,12,13,15 and will examine only RCTs. In addition, we will try to elucidate dosage variables that might affect outcome. These variables include the intensity of ultra-sound applied, the size of the area treated, and the duration of treatment for specific problems.
Randomized controlled trials are widely recognized as the best way of comparing the effectiveness of different treatments.34 We do not believe that RCTs are the only method of obtaining information about an intervention. Many laboratory trials of ultrasound, for example, have demonstrated that it can effect changes consistent with healing.7,3540 In addition, other research methods might demonstrate differences and suggest important research hypotheses. Without applying the rigorous criteria implicit in an RCT, however, our degree of certainty about the outcome of such research is limited. Also, these approaches, although providing information about ultrasound, cannot produce clinically applicable evidence of effectiveness.
Randomized controlled trials, like other research methods, can be biased if done badly, and they often have deficiencies in what they report.34 The discrepancy between what aspects of a study should be reported and what aspects of a study are reported has long been known.41 Readers, however, have to judge a study on the basis of what is reported and available in the public domain.
For this reason, the present review was done in 2 stages. In the first stage, we identified a set of relevant RCTs and investigated their methodological adequacy. The criteria each RCT needed to meet for inclusion in the present study were based on those developed by Sackett et al42 and are shown in Table 1. In the second stage, we analyzed relevant aspects of their content in order for us to judge the clinical effectiveness of therapeutic ultra-sound for treating people with pain and musculoskeletal disorders and for promoting soft tissue healing.
|
| Method |
|---|
|
|
|---|
A total of 35 RCTs of therapeutic ultrasound published in the English language were identified and are listed in Table 2. Both authors of this article independently read the articles describing these trials. The initial reading was done in an effort to ensure that all articles described RCTs that investigated the clinical use of ultrasound for treating people with pain or musculoskeletal injuries or for promoting soft tissue healing. Some studies were excluded at this stage. Two studies43,44 using subjects without impairments in what we would consider laboratory conditions were not analyzed further. The first study involved treatment of a local skin inflammation caused by applying ultraviolet therapy,43 and the second study investigated the effects of ultrasound on the mechanical pain threshold in subjects with no known pathology.44 Neither represents one of the conditions affecting soft tissue or causing pain or functional restrictions for which ultrasound is used clinically. Next, 4 studies4548 using multiple interventions were rejected. The method of using ultrasound in these studies made it difficult for us to distinguish any effects of it from that of other interventions given concurrently. Therefore, these 4 studies were not included in this systematic review. Finally, 2 studies49,50 that duplicated other published results were excluded. Both studies presented subjects and results published in other articles that are included in our review. We then applied the methodological filters shown in Table 1 to the remaining 27 RCTs.
|
We deemed 26 of the 27 remaining studies to be adequate for further analysis. The study that we excluded as not adequate for further analysis had 2 groups, a group that received ultrasound and the standard treatment and a group that received the standard treatment only.18 Without including a group that receives both placebo ultrasound and the standard treatment, we believe that it is impossible to distinguish the placebo component in the outcome possibly contributed by using therapeutic ultrasound equipment.
Filter 2Blinding of Assessors, Subjects, and Users
Three aspects of experimental blinding (methodological filter 2) were considered: (1) blinding of the assessor, (2) blinding of the subjects, and (3) blinding of the users of the ultrasound equipment (ie, the therapists). There was insufficient evidence of adequate blinding on at least one of these grounds for 16 of the remaining 26 studies.16,17,20,21,23,24,26,29,30,32,33,5155
Blinding of subjects and therapists is a complex and important issue when using ultrasound. Intensity is only one factor affecting the ability of subjects and therapists to identify whether there is or is not an output. Other factors that we believe can contribute to an accurate identification of an output are the speed of applicator movement, the anatomical location insonated, and the temperature of the contact medium used.
The articles by Dyson et al16 and Downing and Weinstein52 indicate the extent to which subjects and therapists can correctly identify whether the subjects are receiving ultrasound or a placebo. Dyson et al16 reported that 6 (66.6%) of the 9 subjects in the experimental group detected the skin heating produced by pulsed ultrasound with an estimated space-averaged time-averaged (SATA) intensity of 0.2 W/cm2 using a 1-MHz applicator and a dosage of 1 minute per 0.5 cm2 surface area. Downing and Weinstein52 reported that 50% of the volunteers who were each treated 6 times could correctly identify whether their treatment consisted of active ultrasound or placebo ultrasound. Unfortunately, no information was provided on the intensity used, which we believe is a major factor. Downing and Weinstein52 then used warmed gel to increase the likelihood of subjects not knowing whether the ultrasound machine was producing sound waves. The mean level of ultra-sound dosage was 1.2 to 1.3 W/cm2 for an area of about 150 cm2 for 6 minutes, a higher dosage than is used in many studies but also a larger area insonated. Six (40%) of 20 subjects guessed correctly whether they were receiving active ultrasound or placebo ultrasound, and the therapist correctly identified 8 (72.7%) of the 11 subjects who received ultrasound. These findings indicate the difficulties in ensuring that both subjects and therapists are unaware of the output status of an ultra-sound machine (double blinding).
McLachlan28 decreased the likelihood of both subject and operator identification of active or placebo ultra-sound through local heating by using a resistor to produce some heating in a modified, but otherwise identical, ultrasound applicator. In view of the problems in implementing effective blinding strategies using ultra-sound, no studies could be excluded with certainty because of this methodological problem. As a result, 26 studies were left.
There was, however, one exception to our ignoring the blinding method.29 In a study in which the investigators used an applicator with a discernibly different output (45 kHz), the authors recognized this as a problem due to the frequency of ultrasound used.29 They commented that their method was inadequate for double blinding. All therapists and many patients would have been aware that there was no output with the placebo 45-kHz ultrasound. At that frequency, ultrasound has a barely audible output and produces a detectable cutaneous sensation. Excluding that study29 left us with 25 studies as methodologically acceptable.
Filter 3Treatment Variables
In 9 of the remaining 25 trials, the researchers did not provide sufficient details about treatment to allow replication of the study. In 8 trials, there was no indication of whether the output of the ultrasound machines used was checked.20,21,23,30,35,5557 Several researchers5861 have measured ultrasound machine output and reported that it frequently varied more than 30% of that indicated. We believe that such results justify concerns about possible discrepancies between the dosage displayed on the machine and that given to the patient. Such discrepancies could affect outcome and have been suggested as a possible reason why ultrasound treatments are believed to be ineffective.62 With the exception of those listed, the remaining RCTs documented that the machine's output was checked, in some cases prior to each session.
We excluded one study24 because we believed there were inadequate details of the intensity of ultrasound applied. Of the remaining 16 studies, a frequency of 3 MHz was used in 7 studies (3.28 MHz in the study by ter Riet et al22) and a frequency of 1 MHz was used in 9 studies (0.89 MHz in the study by Ebenbichler et al31 and 1.1 MHz in the study by McLachlan28). (McLachlan28 provided the name of the machine used but not its frequency; the Medtron P3001,* used in that study has a frequency of 1.1 MHz.) In the study by Hashish et al,25 the SATA intensities for the 3-MHz machines ranged from 0.02 to 0.3 W/cm2. The SATA intensities for the 1-MHz machines ranged from 0.2 W/cm2 in 3 RCTs6,54,63 to 2.4 to 2.6 W/cm2 in 2 RCTs.28,53 Researchers who commented on their selection of treatment variables said that their choice reflected practice.22,28
Details of either the effective radiating area (ERA) or the geometric area of the applicator were accepted for the purpose of our study. Because the piezoelectric element that generates the ultrasound does not vibrate uniformly, the ERA of the applicator is smaller than its geometric area.64 As a consequence, calculations of dosage based on geometric area may, in some cases, slightly understate the actual wattage per unit area applied relative to those using ERA. As the applicator needs to be kept moving to avoid hot spots, the area affected by ultrasound energy cannot be precisely determined and the dosage at a particular depth of tissue cannot be known.64,65 The beam nonuniformity ratio (BNR) expresses one variable contributing to this outcome; others include the wattage applied, the depth and types of tissue, and the frequency of ultrasound. The BNR is not relevant to our article, given the omissions of more basic aspects of dosage, but it will have to be considered if appropriate dosages are ever to be determined through research.
The area treated and the duration of application also affect dosage. All research reports that we reviewed contained information on the treatment duration but often not the size of the area treated. Although constant movement of the applicator means underlying tissues receive a variety of ultrasound energy, we believe that details of the size of area treated (ideally volume) should always be provided by researchers. The studies we exam-ined provided descriptions of the areas insonated, enabling an estimation of the size of area treated.
Filter 4Outcome Measures
We believed that the different measures used for patient outcomes (methodological filter 4) in the remaining 16 studies were acceptable. In each study we judged, at least one outcome measure with face validity was used. We also believed that these measures have a widely and generally acceptable level of reliability, as evidenced by their use in more than one study. For example, in the studies in which the effect of ultrasound on ulcers was examined, there was use of tracings or a similar method of measuring ulcer area.19,22 In some studies,32,51,53 the measures used included the scoring of pain using a visual analog scale. Where appropriate, additional measures such as grip strength were used.51,63
Some authors used what we would consider unacceptable measures; however, because multiple outcome measures were used, these studies were not excluded from our study. For example, Ebenlicher et al54 used an additive scale for ordinal data, but without any evidence of the appropriate Rasch analysis that we believe is necessary to justify that approach. Although we believed that particular measure not valid, other measures they used, in our opinion, were acceptable.
Filter 5Sample Size and Power
We believe that sample size is only an issue when there are negative findings (methodological filter 5). A power analysis indicated that 3 of the RCTs19,32,52 that remained following screening had results that indicated that ultra-sound is no more effective than placebo treatment, but we believe that these RCTs had too few subjects to identify even a large treatment effect. Following dropouts, Gam et al32 had 18 subjects in the active ultra-sound, massage, and exercise group; 22 subjects in the placebo ultrasound, massage, and exercise group; and 18 untreated control subjects. Downing and Weinstein52 treated all 20 subjects in their study with active, active assisted, and passive range of movement exercises, followed by active ultrasound given to 11 subjects and placebo ultrasound given to 9 subjects. McDiarmid et al19 had 21 subjects in their active ultrasound treatment group and 19 subjects in the placebo ultrasound group. To attain an 80% probability of being able to detect a treatment effect (alpha=.05) requires a minimum of 26 subjects per group for a 2-group study if a large difference in outcome is expected and a parametric statistic is used.66 If a nonparametric statistic is used, as in these studies, up to 20% more subjects are needed if a large treatment effect is to be identified.66 This methodological filter of sample size left us with 8 studies in which there were no discernable treatment effects and with 5 studies with positive findings.
Filter 6Data Analysis
We identified 3 of the 13 remaining studies as having problems with aspects of their data analysis.16,17,63 The analysis of results presented by Dyson et al16 is confusing because the number of subjects (ulcers or patients) in each group in the first part of their study is unclear. There is another major problem of ulcer variability in that study. The authors noted that the chronic varicose ulcers tested ranged in size from 1.5 to 12.75 cm2 and had existed from 6 to 360 months in their subjects and that the response to intervention was very variable. Although robust, a Student t test, like other parametric tests, is dependent on a level of homogeneity of variance.66 We believe that this is especially important when small numbers of subjects are used and, in this context, suggests that a nonparametric analysis rather than a parametric analysis should have been used. Either problem, we believed, was sufficient to exclude this study from further consideration.
Roche and West17 and Binder et al63 examined the effects of ultrasound, but they did not demonstrate an equivalence of the experimental and control subject groups before the study began. In the study by Roche and West,17 the ulcer size was markedly less in the placebo ultrasound group (23.62 versus 32.51 cm2) and of longer duration (12.35 versus 5.37 years) than for the active ultrasound group. Binder et al63 provided evidence of equivalence of patient details such as age but not of the initial measures of grip strength or of pain used to evaluate the effect of using ultrasound. Consequently, any apparent differences between the groups after treatment may have been due to differences between subjects in each group. For example, more subjects in the ultrasound group may have had lesions that were likely to recover more quickly, perhaps because of their relative recency or lesser severity or because of other factors such as the age and activity level of a subject.
Initial equivalence of groups cannot be assumed; therefore, we believe the results are not compelling. Dyson62 identified another problem in the study by Binder et al.63 Subjects with poor results from treatment typically did not rest adequately from the precipitating cause. The etiology of overuse injury suggests that rest is likely to be effective in assisting the resolution of this condition.62 We found, however, that the study by Binder et al63 provided too few details to examine this claim and to differentiate the contributions of active ultrasound from those of resting in patients with lateral epicondylitis.
Summary
From the original 35 studies identified, Table 2 shows that 10 studies remained.6,22,2528,31,51,53,54 In 2 of those 10 studies, there were differences in outcomes as those subjects treated with active ultrasound improved but those subjects treated with placebo ultrasound did not improve.31,54 In the other 8 studies, no differences between groups treated with active ultrasound or placebo ultrasound were found.
| Analysis of Methodologically Acceptable Studies |
|---|
|
|
|---|
Dosage
Table 3 shows the conditions that the subjects had and the dosages used in the 10 studies. Some details had to be estimated, specifically, the size of the applicator for articles published prior to 19922528,51 and the size of area treated.6,2528,31,54 For the applicator, this is likely to be correct as machines used prior to 1992 typically had 5-cm2 applicators with a smaller ERA. In the absence of information on the geometric size of the applicator, we assumed it to equal the ERA. This produces, in some instances, a small overestimation of the wattage applied. With the size of area treated, this was estimated from descriptions of the area treated and is clearly a potential source of error when used in subsequent calculations of energy density.6,2528,31,54
|
The energy density (total energy [in joules] per unit area [in square centimeters]) for the remaining studies was from a low of 2 J/cm2 with a frequency of 3 MHz in the study by Hashish et al25 to a high of 150 J/cm2 with a frequency of 1 MHz in the study by Ebenbichler et al31 (
=55.79 J/cm2, 95% confidence interval=19.884.3). This is a large range. For example, in the 2 studies in which shoulders were treated,6,31 a 1-MHz frequency and energy densities of 40 and 150 J/cm2, respectively, were used. Carpal tunnel syndrome and epicondylalgia were treated using a frequency of 1 MHz and energy densities of 60 and 120 J/cm2 in the studies by Ebenbichler et al54 and Lundeberg et al51, respectively. These differences in the energy density used suggest that comparable areas are not treated with similar dosages. There also was no apparent relationship between the year of the study and energy density applied. In the first study51 and the most recent study31 using a 1-MHz frequency, researchers applied the highest energy densities. Table 3 shows that a similar phenomenon occurred among studies using a 3-MHz frequency.
The limitation of this analysis was the uncertainty added by the use of some estimates in the calculations. We based estimates we made of the geometric size of the applicator2527,51 and the size of the area treated6,2528,31,54 on the details provided in the respective articles. We made these estimates to compare the effect of different levels of dosage (energy density) on patient outcomes. The data, however, suggest that there was considerable variation in energy density that is not accounted for by the type of patient problem, by the size or depth of the area treated, or by the year of the study.
Problems Treated
The diversity of problems treated with ultrasound limits comparisons between studies and possible conclusions on effective dosages. Each study reviewed compared the effects of active and placebo ultrasound. Depending on the condition treated, both groups also had either identical concurrent treatment or no additional concurrent treatment. This allowed the contribution of ultra-sound to be distinguished from other components of multiple interventions, the aim of an RCT. Between-study comparisons were difficult even for studies that seemed similar because of differences in the inclusion criteria used. For example, Ebenlicher et al31 used ultrasound for people with calcific tendinitis of the shoulder, whereas Nykanen6 simply said that the subjects had shoulder pain. The possibility of a meaningful comparison is limited by one study6 having included a wider range of shoulder problems than the other study.31
Outcome Measures Used
A variety of outcome measures also were used. At least one study54 had a problematic outcome measure, discussed earlier, in which ordinal data were added. In that study, however, the authors also examined the results of electroneurological testing, which was used as a diagnostic tool and as an outcome measure. Nerve conduction studies can provide an accurate assessment of conditions involving compromise of peripheral nerves,67,68 including carpal tunnel syndrome.69 Because of this, we judged the study as acceptable, as were the other remaining studies. We contend, therefore, that differences in findings were not attributable to problems with outcome measures.
Outcomes for True Control Groups
In all 10 of the studies, active ultrasound and placebo ultrasound were used. In 3 studies25,26,51 there was what we would consider a true control group, which received neither active ultrasound nor placebo ultrasound. In each of these studies, the group that received active ultrasound had better outcomes than the true control group (outcome measures: pain,25,26,51 facial swelling,25,26 trismus,25,26 serum C-reactive proteins,25,26 plasma cortisol,26 weight test,51 pain and power with wrist dorsiflexion,51 and grip strength test51). In the 2 studies in which the group that received the placebo treatment and true control group were compared, the authors reported better outcomes for the group that received the placebo treatment than for the true control group.25,26 This finding provides some support for the use of ultrasound equipment for treating some conditions.28 Such a finding is also consistent with known responses to pain, an outcome measure in many of the trials we examined.
Differences Between 2 Categories of Studies
We identified 2 categories of methodologically acceptable studies: studies in which ultrasound was found to produce desirable outcomes and studies in which that was not found. The same first authors published the only 2 studies showing active ultrasound as more effective than placebo ultrasound that were, we believe, methodologically adequate.31,54 This raises obvious questions as to how those studies differ from the other 8 studies.
Not all methodological filters were strictly applied. In particular, whether experimental blinding (filter 2) was effectively implemented during ultrasound use was not certain in 16 of the 26 studies remaining at that stage in the screening process. This issue is relevant to both studies that showed active ultrasound had desirable outcomes. Blinding of subjects and therapists was not addressed in the earlier study by Ebenbichler and colleagues.54 In the later study by Ebenbichler and colleagues,31 subjects and therapists were said to be unaware of which treatment was the placebo ultrasound treatment and which treatment was the active ultrasound treatment, but no details were provided. Had filter 2 been strictly applied, both studies would have been excluded early in the screening process.
In both studies by Ebenbichler and colleagues,31,54 the investigators applied ultrasound when it was possibly not the ideal treatment. In response to a critical letter about one of the studies,70 Ebenbichler described the purpose in their study as "investigating the efficacy of a promising entityultrasound treatment."71 He agreed with the letter writer that there possibly was a relatively high relapse rate in patients with carpal tunnel syndrome treated with ultrasound. In a letter commenting on a different study investigating the use of ultrasound in people with carpal tunnel syndrome, a physiatrist contended that there were risks of aggravating the condition by heating directly over the nerve rather than over associated structures.72 This, however, does not detract from the fact that Ebenbichler et al54 reported a difference in outcome for those treated with ultrasound (outcome measures provided: subjective symptom score for main complaint and sensory loss; electroneuro-graphic measurements of median motor nerve distal latency, antidromic sensory nerve conduction rate, and peak-to-peak amplitude; and physical function levels, including strength of handgrip and of finger pinch).
In the later study by Ebenbichler et al,31 the researchers treated individuals with calcific tendinitis of the shoulder. A feature of this condition can be the spontaneous resorption of the calcium deposits over time.48 We excluded one RCT48 from our analysis because it compared no treatment with a treatment of ultrasound and acetic acid iontophoresis. The study by Perron and Malouin48 had 2 groupsa treatment group and a true control group. The authors found no differences between the 2 groups over the 3 weeks of the study. During this time, both groups had positive changes, with a decreased size and density of calcium deposits and an increased range of passive shoulder abduction with decreased pain during it. In no instance was the change in any outcome measure statistically significantly different between the groups. That is, the natural course of calcific tendinitis was unaltered by ultrasound.
| Discussion |
|---|
|
|
|---|
Ultrasound has been used therapeutically for over 6 decades in the ways reported in the trials examined in this study.73 Any clinically significant effects should, by now, have been identified in a number of rigorous studies that showed which patient outcomes are improved by using therapeutic ultrasound. In our review, we found that is not the case. Furthermore, in the few methodologically adequate studies that exist, treatment was provided for a wide range of problems; thus, few conclusions can be drawn. Similarly, no replications exist of studies with significant findings. Having different researchers in a different facility using the same procedure and obtaining a similar finding would considerably affect the strength of our certainty about ultrasound.
We found that the dosages of ultrasound used in the studies we reviewed varied considerably and for reasons that were not always clear. No underlying patterns were evident except possibly that the studies with significant outcomes were among those using a higher total energy output. Furthermore, without adequate data, there is little scientific basis for dosage selection in clinical practice. This leaves a question of the extent to which the diversity of dosages used helps explain the limited evidence of effectiveness of therapeutic ultrasound.
Limitations
One possible limitation of our review is its exclusive focus on RCTs. There are a number of other methods of obtaining relevant information about therapeutic ultra-sound. Many laboratory studies indicate that ultrasound has an in vitro effect (see "A Review of Therapeutic Ultrasound: Biophysical Effects" by Baker et al in this issue). Unless these effects are not only consistent with healing but also sufficient to alter a relevant patient outcome positively, they do not justify the clinical use of ultrasound. Based on their clinical experiences, many therapists believe that ultrasound benefits healing. Until methodologically adequate studies can demonstrate that people treated with active ultrasound consistently have a better outcome than those treated with placebo ultra-sound, we believe that doubts must remain.
Another possible limitation is the particular criteria used as filters on the set of studies identified. Those criteria, however, are entirely consistent with other sets used in previous systematic reviews and meta-analyses.5,12,15 In addition, they are consistent with requirements given for reporting RCTs described in major medical journals published in the United States41 and the United Kingdom.34
One possibly valid criticism of our review is that we did not apply the different filters with sufficient rigor. For example, had studies that provided dubious or imprecise details of blinding of subjects, assessors, and therapists been excluded, few would have passed filter 2. Similarly, had the criterion regarding the establishment of controls (filter 1) been rigorously applied, few studies would have passed, as many authors did not provide adequate details of how they randomly assigned subjects to groups. Consistent with the possibility of a differential application of filters, there are some differences between this review and the review by van der Windt et al.15 They accepted 2 studies52,56 that we rejected as methodologically inadequate. Had we accepted those 2 studies, however, it would have made no difference to the outcome of this review, because both studies demonstrated no differences when using active ultrasound rather than placebo ultrasound.
| Conclusions |
|---|
|
|
|---|
| Future Directions |
|---|
|
|
|---|
| Footnotes |
|---|
* Medtron Medical, 57 Aster Ave, Carrum Downs, Victoria 3201, Australia. ![]()
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J Taradaj, A Franek, L Brzezinska-Wcislo, L Cierpka, P Dolibog, D Chmielewska, E Blaszczak, and D Kusz The use of therapeutic ultrasound in venous leg ulcers: a randomized, controlled clinical trial Phlebology, August 1, 2008; 23(4): 178 - 183. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. J Robertson Invited Commentary Physical Therapy, January 1, 2008; 88(1): 58 - 61. [Full Text] [PDF] |
||||
![]() |
S. J Straub, L. D Johns, and S. M Howard Author Response Physical Therapy, January 1, 2008; 88(1): 61 - 62. [Full Text] [PDF] |
||||
![]() |
S. J Straub, L. D Johns, and S. M Howard Variability in Effective Radiating Area at 1 MHz Affects Ultrasound Treatment Intensity Physical Therapy, January 1, 2008; 88(1): 50 - 57. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. J Robertson Invited Commentary Physical Therapy, August 1, 2007; 87(8): 995 - 999. [Full Text] [PDF] |
||||
![]() |
R. A Wong, B. Schumann, R. Townsend, and C. A Phelps Author Response Physical Therapy, August 1, 2007; 87(8): 999 - 1001. [Full Text] [PDF] |
||||
![]() |
R. A Wong, B. Schumann, R. Townsend, and C. A Phelps A Survey |